Light Source: 6V 10W, halogen lamp of more than 2000 hours’ life time
Spectrometer: interference filter.
Wavelength: 340nm、380nm、405nm、505nm、546nm、578nm、620nm, and one wavelength free.
CML precision: ±2nm
FWHM: ≤10nm
Stray light:≤0.5%(absorbance≥2.5)
Absorbance range: -0.214~3.000
Absorbance linearity: ±1%
Absorbance repeatability: ±0.001
Absorbance resolution: 0.0001
Precision: ≤1.0﹪
Carry over: ≤1%
Temperature: 25℃、30℃、37℃,±0.3℃
Flow cell:permanent quartz flow cell, minimal volume is 32μL, and disposable flow cell is selectable
Sampling system: pressure auto releasing pumps ensure the accuracy of suction volume
Operational system: VISTA、Windows XP、Windows2000
Test Method: End point,Fixed time,Kinetic,Factor,Muli-standards,Bichromatic,Dual wavelengths
Parameter Setting: Patients’ profile and calculated item can be set easily
Absoption volume: 200μL~3000μL
Sample volume: 10μL~50μL
Reaction curve: real time display the reaction curve
Memory: Able to store more than 200 parameters while 55 parameters of them are stablized ;
Able to save more than 15000 test results and auto save the results if the power is off during working
QC: auto draw the QC figure
Reagent: Open reagent
Display: LCD with high brightness
Input: key button and touch screen
Printer: Inner thermal printer,able to be connected to external printer